Table 9 Various artefacts in EEG signal recording

From: Mental state and emotion detection from musically stimulated EEG

Category

Artefact /Source(Cause)/Frequency/Amplitude Morphology

Artefact Prevention

Physiological

Artefacts

Cardiac/Heart/\(\le\)1Hz/1-10mV

/Epilepsy

Selection of proper montage

Monitoring during recording

Offline visual inspection

Low pass filter (LPF)

Data Rejection

EOG/Eye/0.5-3 Hz/100mV

/Tumour, delta wave

Artefact-free recording protocol

Online monitoring

Offline visual inspection

Low pass filter (LPF)

Using various

Transform (ICA, PCA, EOG subtraction)

Muscle Artefact/Muscle

/\(\ge\) 100Hz/low /Beta frequency

Artefact-free recording protocol

Online Monitoring

Offline visual inspection

High pass filter(HPF)

Data rejection

Physical movement artefact

/Physical movement

/Very low/ very high

/Morphology different

from actual EEG

Artefact-free recording protocol

Online monitoring

Offline visual inspection

Data rejection

External

Artefacts

Transmission line

/Transmission line

50–60 Hz/low/

Morphology different

from actual EEG

Notch filter

DC power supply

Phone artefacts

/Mobile and landline phone

/high/high/different

Artefact-free recording protocol

Electrode artefact

/Electrode and sweating

/very low/high

Artefact-free recording protocol

LPF

Impedance artefact

/Electrode with impedance>5K\(\Omega\)

/-/approximately 100 μV/different